Meriam M130 T_C Thermocouple Calibrator User Manual

Page 16

Advertising
background image

M130 – T/C Calibrator

Instruction Manual Z1017-IR

E.O. 6566 Sept. 2008

www.meriam.com

Page 16 of 21

Specifications

Type and Range:

T/C types B, E, J, K, N, R, S, T & milli-volts
Milli-volts range -15mV to +80.00 mV

T/C Measurement & Simulation Accuracy: See Table below
mV Accuracy: 0.005% of reading ±6µV
mV Temp. Effect: ±2 µV /

°C (Factory Ref. Temp. = 23°C)

Cold Junction Effect: ±0.05

°C/°C

Cold Junction Accuracy: ±0.1

°C at 25°C

T/C Resolution:

T/C types E, J, K, N, T: 0.1

°C

T/C types B, R, S: 1

°C

mV: 0.001 mV (1 µV)

Temperature:

Storage = -40

°C to +85°C (-40°F to +185°F)

Operating = -10

°C to +50°C (14°F to +122°F)

Humidity: 10% to 90% Rh
Connection:
Measure: standard mini-T/C connector
Simulate: standard mini-T/C connector
Zero: supplied mini-T/C connector with internal copper short
Maximum Input Voltage: 30VDC
Battery Type:
3 x AA battery, field replaceable.
Battery Operation:

>100

hours continuous use without the backlight for measure

or simulation modes, 1 year shelf life, low battery warning at
approximately 2 hours remaining run time.

Enclosure:

(6.5” Ч 3.2” Ч 1.1”) Polycarbonate, Permanently Static
Dissipative, ESD Protection

Enclosure with Boot:

(6.8” Ч 3.5” Ч 1.3”)









Advertising