Multichannel Systems STG1000 Manual User Manual

Page 72

Advertising
background image

Stimulus Generator 1000 Series User Manual

66

Fig. 11 STG output with programmed biphasic 200 mV and 20 μA pulses, duration 100 μs for

each phase.

With a voltage amplitude of 200 mV (U, yellow), there is almost no delay between the positive
and negative phase. The top part of the figure shows all three voltage steps (50 mV, 100 mV, and
200 mV). You can clearly see that the delay between the phases decreases significantly with
increasing pulse amplitude.

Advertising