Multichannel Systems STG1000 Manual User Manual

Page 80

Advertising
background image

Stimulus Generator 1000 Series User Manual

74

Operating system

Windows

© 98, ME, NT, 2000, or XP; English and German versions

supported

Data import

ASCII file format

13.2 Technical Specifications STG1004/8

Operating temperature

10 °C to 35 °C

Storage temperature

0 °C to 50 °C

Dimensions (W x D x H)

275 mm x 330 mm x 115 mm

Weight 4.8

kg

Fuse

0.2 A @ 230 V, slow blow; 0.4 A @ 115 V, slow blow

Supply voltage

Available types: 100 VAC, 115 VAC, or 230 VAC @ 50 to 60 Hz

Number of analog output
channels

4 (STG1004) or 8 (STG1008)

Output voltage

8 V to + 8V @ max. +/-20 mA

Output voltage resolution

2 mV

Ouput voltage slope

> 4 V/μs

Rise time (10-90 %)

3 to 5 μs @ delta U = 1 V or @ amplitude > 100 μA & R

L

= 10

kilohms

Output current

0.8 mA to +0.8 mA @ max. 100 V compliance voltage (between
+I and I)

Output current resolution

200 nA

Output current slope

100 μA/μs @ R

L

= 10 kilohms

Time lag between Sync Out
and voltage output

12 μs +/-2 @ amplitude > 200 mV

Time lag between Sync Out
and current output

15 μs +/-2 @ amplitude > 100 μA & R

L

= 10 kilohms)

Resolution

13 bit

Time resolution

20 μs

Output signals

Freely programmable (rectangular, ramp, sinusoidal)

Maximum frequency
(rectangular waveform)

25 kHz

Timing on external trigger

Minimum time length of
trigger input (TTL)

20 μs

Time lag between trigger
input and stimulus output

120 +/ 10 μs

Time lag between trigger
input and Sync Out

110 +/ 10 μs

Maximum trigger input
frequency

50 Hz

Interface and connectors

Sync Out for synchronization
with following devices

Digital channel via coax cable

Trigger In for synchronization
with preceding devices

Digital channel via coax cable

Interface (connection to
computer)

RS232 serial port

Download rate

56 Kbps

Advertising