Multichannel Systems STG3008-FA Manual User Manual

Page 46

Advertising
background image

STG3008-FA Manual

40

The following picture shows a file of format type 6 with header.

Using the format type 6 without header, only the first channel of the MC_Rack raw data file can
be imported.

Using the format type 6 with header you choose how many channels are imported in "channels",
line 3. The output mode in line 4 has to be voltage. You choose which channel should be
imported first in "channels" line 6. You can add the channels you also want to import one after
the other in any order.

The value microvolt is normally used for data recording in voltage mode of MC_Rack. The STG
generates voltage pulses in values of millivolt. Please use the “Amplification” factor of the ASCII
Import Dialog to convert μV in mV, otherwise the output of the STG is zero.

The amplification factor is generally applied in files with and without header.

ASCII Import of Files

For ASCII import of files, please use the command ASCII Import in main menu "Settings".

Advertising