STG3008-FA Manual

58

Operating the STG in Batch mode

Important: You can operate the STG in

Batch

mode only with a

valid computer connection

.

A file in the batch list is only downloaded

after

the preceding file has been run.

1. On

the

Batch menu, click Open to open a batch file.

The Open dialog box appears.

2. Browse your folders and select a valid batch file, for example the sample file TetanusBatch.stb

in the MC_Stimulus II program folder.

3. The

Batch Status Display opens. It lists all stimulus files in the batch and the duration

of stimulation.

4. On

the

Batch menu, click Start to start the download of the first file in the batch list and the STG

if the option Start STG after Batch Download on the Settings menu is selected. The progress
of the pulse protocol is displayed in the Status column of the Batch Status Display.

5. Click

Stop on the Batch menu to stop the batch if necessary.

6. Close

the

Batch Status Display with the Close command.

Warning: The STG starts immediately after the download if the option

Start STG after Batch

Download

on the

Settings

menu has been selected. Make sure to deselect this option if you

do not want to start the stimulation right away.

Autorepeat option

If you select the Autorepeat Batch Mode option on the Settings menu, the batch is restarted
with the first file after the last file in the batch has been finished.