Multichannel Systems STG3008-FA Manual User Manual

Page 70

Advertising
background image

STG3008-FA Manual

64

Resistance dependency of rise time

The higher the load resistance, the higher will be the rise time. The rise time (10 % / 90 %)
was measured with increasing load resistances and a constant current amplitude of 50 μA.
For 50 μA pulses, the relationship is linear up to the maximum load resistance (please see
"Current Mode") as shown by the following curve. Not shown are the results of the same
measurement under the same conditions, except of the amplitude with a constant current
amplitude of 100 μA. Up to a load resistance of 450 k

, the results are congruent with 50 μA

amplitude.

Stimulus output comparison

The following figures are examples for demonstrating how the amplitude and resistance will
affect the kinetics of a current pulse. These images document also that in contrast to the
current mode, in voltage mode, the rise time is constant and not dependent on the amplitude
or load resistance.

The graphs were generated by a standard oscilloscope. The devices were set up according to the
recommendations under "Voltage Mode" and "Current Mode". The same pulse protocol and the
same channel number of the STG outputs were used for measuring the voltage and the current
output.