Installation and Operation

11

Example:

You are using a MEA1060-UP amplifier for upright microscopes on channel 1, and a perfusion
cannula PH01 on channel 2 of a TC02. You have to configure each channel for the appropriate
instrument. Select, for example, MEA2100 for channel 1 and PH01 for channel 2 in the
Channel Configuration menu of the TC02.

Note: The factory default parameters were optimized for an ambient temperature. The
configuration for use with the PH01 were optimized for a medium flow rate. Under extreme
conditions, you may have to adjust the configuration for your experimental setup.

3.3.6 Hardware

Diagnosis

This menu should be used for reviewing parameter settings or check the hardware performance
in case that you observe any trouble with the instrument. Each channel can be checked separately.
If the problem persists, please contact your local retailer. The highly qualified staff will be glad to
help you. Keep the displayed information at hand when contacting customer support.

There are four separate screen views with different sets of information in the Diagnosis menu.
You can toggle between the views by pressing the Up and Down buttons.

Diagnosis 1: Measured values

This diagnose screen view is used for checking the temperature sensor.

Temperature

Actual temperature

Resistance 2

Cable resistance high side of sensor, see also chapter
"D-Sub9 Pin Assignment".

Resistance 1

Cable resistance low side of sensor, see also chapter
"D-Sub9 Pin Assignment".

Resistance X

Sensor resistance plus cable resistance

Resistance S

Sensor resistance

Board Temp

Board temperature (The TC01/02 will switch off the channel outputs
and go into stand-by mode when the board temperature reaches 90 °C.)