4.3 Parameter

Ranges

The setpoint temperature and the PI coefficients can be modified in the following ranges.
The maximum power of the TCX is 30 W. If you connect a device with a maximum power less
than 30 W, please decrease the maximum power to protect the device against destruction.

Parameter

Range

T

0.0 to 105.0

P

0.1 to 99.99

I

0.01 to 100.0

Power

0 to 30 W

4.4 MCS Default PI Coefficients

Note: The following PI parameters have been optimized at an ambient temperature of 25 °C,
the PI coefficients for use with the PH01 at a flow rate of 3 ml/min. You may have to adjust these
PI coefficients for your experimental setup, especially if the ambient temperature or the flow rate
differ at large from those used by MCS. Using suboptimal PI coefficients can lead to an oscillation
of the actual temperature, which is harmless, but may result in an unwanted behavior of the
temperature controller.

OP

Table TCW PH01

MEA1060-UP MEA1060-INV USB-MEA256 MEA2100

P 10.00

60.00 2.00 4.00

6.00

6.00

6.00

I 1.00

75.00 2.00 1.50

0.90

0.90

0.90