Installation PEM-9004

Teledyne Analytical Instruments

14

Connect the gas out port to a suitable vent.

WARNING: TOXIC AND FLAMMABLE GASES ARE PRESENT IN

THIS INSTRUMENT. THESE GASES CAN BE
HARMFUL AND/OR FLAMMABLE EVEN AT LOW
CONCENTRATIONS. THE ANALYZER’S EXHAUST
PORT MUST BE VENTED TO A SAFE AREA.
PERSONAL PROTECTIVE EQUIPMENT MUST BE
WORN WHEN USING OR SERVICING THIS
INSTRUMENT.

AFTER EACH USE, BEFORE TURNING OFF THE
POWER OR DISCONNECTING ANY TUBING
CONNECTED TO THE ANALYZER, REMOVE THE
PROBE AND FLUSH THE UNIT WITH AIR FOR FIVE
MINUTES

Figure 3-1: PEM-9004 Front Panel

3.3 Initial Calibration

The instrument, when first powered up will go through a self-

diagnostic test routine and auto calibrate. For this reason, when