Portable Emissions Analyzer

Operation

Teledyne

Analytical

Instruments

55

U batt. = current voltage

I Bat = current amperage

T Bat = measured battery temperature

Cap. = current battery capacity

To perform a measurement while charging, press ENTER from the

Charge Control menu. While in the measurement mode the battery will
be recharging and continuously monitored by the system.

When the battery is fully charged, the analyzer automatically

switches to the passive recharging mode (trickle charging) and the
charge control screen disappears.

When active recharging is finished the charger can remain

connected to the device without damaging the battery.

CAUTION:

USE ONLY THE SUPPLIED OR FACTORY
AUTHORIZED

CHARGER

UNIT FOR RECHARGING

THE BATTERIES. ANY OTHER CHARGER MAY
CAUSE

DAMAGE TO THE BATTERY AND/OR

THE

ANALYZER.

4.11 Battery / Line Voltage Operation

B a t t e r y o p e r a t i o n :

Maximum of 36 hours of continuous
measuring (with backlight).

B a t t e r y c h a r g e r :

95-135 VAC 60 Hz USA version
200-250 VAC European version

The analyzer employs Intelligent monitoring by means of an

integrated charge-management-system. To maintain the service life and
performance of the NiCd battery, please observe the following battery
charging instructions.

4.11.1 Charging t h e Battery

The PEM-9004 is equipped with an NiCd storage battery. The

service life and capacity of the battery are considerably affected by the
way the instrument is charged and used. In order to make the handling