Operation PEM-9004

Teledyne Analytical Instruments

48

SO

2

and NO

2

gases have a high solubility in water. For

measurement of SO

2

and NO

2

concentrations it is necessary to remove

the condensate residues form the gas filtration and drying system. These
residues can absorb SO

2

and NO

2

which could cause erroneous

measurements.

Furthermore, when carrying out SO

2

and NO

2

relevant

measurements no additional desiccant should be used. Even when dry
the dessicant can absorb a significant amount of SO

2

and NO

2

.

4.6 Macro Start

The Macro Start function launches configuration settings

previously programmed and saved as a macro. See Section 4.5.1.10.

Macros greatly facilitate the efficient operation of the PEM-9004

by loading and initiating convenient configuration settings using only a
few keystrokes. Up to three customized macros can be programmed and
used.

To launch a previously saved macro call the Macro Start function

by pressing the cycle button from the Main menu.