Teledyne AX300-I - Medical application oxygen analyzer User Manual

Page 12

Advertising
background image

Introduction AX300-I

Teledyne

Analytical

Instruments

4

The diverter is not necessary, and should not be used, when the

sensor is placed directly in a chamber or when the sensor is used in con-
fined volume analysis such as incubators and inhalation tents.

CAUTION STATEMENTS:

THE R17MED SENSOR CONTAINS A CAUSTIC ELECTRO-
LYTE AND LEAD. DO NOT TRY TO OPEN THE SENSOR
ASSEMBLY. CHECK THE SENSOR PERIODICALLY FOR
LEAKS. IF THE SENSOR IS LEAKING, REPLACE IT. DO
NOT TRY TO REPAIR IT. CONTACT YOUR DISTRIBUTOR
OR TELEDYNE FOR THE MATERIAL SAFETY DATA SHEET
RELATED TO HANDLING AND DISPOSAL OF THE SEN-
SOR.

THE R17MED SENSOR CONTAINS A CAUSTIC ELECTRO-
LYTE, AND IN THE EVENT OF A LEAK, THE ELECTROLYTE
MAY ATTACK CERTAIN MATERIALS IT IS EXPOSED TO,
SUCH AS ALUMINUM. USE PRECAUTION IN YOUR
CHOICE OF INTERFACE MATERIALS.

REMOVE AND SAVE THE DIVERTER WHEN THE SENSOR
IS USED IN CONFINED VOLUME APPLICATIONS.

THE AX300-I OXYGEN ANALYZER, R17MED OXYGEN
SENSOR AND ASSOCIATED HARDWARE ARE INTENDED
FOR USE AS SECONDARY DEVICES TO VERIFY THE
CONCENTRATION OF OXYGEN IN GAS MIXTURES. DO
NOT USE THESE DEVICES AS PRIMARY LIFE-SUPPORT
OR PRIMARY MONITORING DEVICES.

THE INFORMATION OBTAINED FROM THE AX300-I
SHOULD NEVER BE USED TO ADJUST A LIFE-SUPPORT
SYSTEM. IT SHOULD ONLY BE USED AS AN INDICATION
THAT THE LIFE SUPPORT SYSTEM OR DEVICE IN THE
BREATHING SYSTEM THAT AX300-I IS BEING USED WITH
MAY REQUIRE SERVICE AND/OR CALIBRATION.