Operating Instructions

AX300-I

Teledyne

Analytical

Instruments

18

Read and understand the manual before using the analyzer

Calibrate every 8 hours or before every use.

Visually inspect the sensor for leakage before each use.

Calibrate using 100% oxygen and check in air.

Test batteries regularly and replace when battery indicator shows low
battery (no bars)

Keep the analyzer, sensor and connections dry, or on the dry side of
the breathing system.

Recalibrate after replacing the batteries.

Recalibrate after replacing the sensor.

Use only alkaline batteries.

Ensure that the R17MED sensor is properly connected to the analyz-
er.

Always use the plastic flow diverter when using the tee adapter.

Remove and save the plastic flow diverter when using the sensor in
non-flowing applications

Use only isopropyl alcohol or mild detergent, if cleaning is required,

– DON’T –

Use this analyzer if you suspect any malfunction.

Use the analyzer in the presence of flammable gases.

Use anything but alkaline batteries.

Autoclave or freeze the sensor or analyzer.

Open or try to repair a leaking or broken sensor.

Immerse the unit or sensor in any liquid.

Pass hot or cold gas mixtures over the sensor.

Expose the unit to devices that produce high levels of radio, short
wave, microwave, x-ray, or high frequency interference.

Use cleaning agents or liquids in the cable receptacles or around the
battery compartment.

Place the unit in a water vapor-saturated environment.

Expose the LCD to excessive sunlight.

Expose the unit to a condensing water environment such as a mist
tent.