Teledyne AX300-I - Medical application oxygen analyzer User Manual

Page 29

Advertising
background image

Portable Oxygen Analyzer

Service Instructions

Teledyne

Analytical

Instruments

21

Symptom

Why

What To Do

Sensor will not read be-
low 22 % after calibration
in 100% O2.

Calibration in 100%
was invalid or the
room air enriched with
excess oxygen.

A) Recalibrate using dry gas
B) Make sure that at least 6"

(30 cm.) of tubing is attached
to the exhaust side of the tee
adapter to prevent back fill-
ing. O2 flow rate should not
exceed 5 l/min.

C) Oxygen concentration at the

sensor is significantly higher
than 21%. Take the analyz-
er to a well-ventilated area
and check the reading
again.

D) Try calibrating with a known

good sensor, and if this fails,
contact customer support.

The sensor does not
react to changes in
oxygen concentration,
or the readings are un-
stable and drifting.

Water may have con-
densed on the sens-
ing surface.
Electrical interference
is disrupting the elec-
tronics

A) Remove the sensor from tee

adapter and unscrew the plas-
tic flow diverter. Using absor-
bent tissue or cotton swab,
gently wipe off sensing sur-
face inside threaded portion of
sensor assembly. Do not
damage the sensing surface,
and do not leave any tis-
sue/swab residue on the
sensing membrane.

B) Relocate unit away from

sources of electrical noise
such as cauterizing equipment
and two-way radios.

The display is flashing

√ SENSOR

The unit has detected
a fault in the signal
from the sensor.
Sensor has expired.
The sensor has been
exposed to a gas con-
taining less than 18%
oxygen.

A) Check sensor cable con-

nections to ensure they are
completely inserted into the
mating connector and the
capture nut is firmly in
place.

B) Expose the sensor in 100%

O2 and check calibration.

C) The sensor output has fal-

len to a level where it is no
longer usable. Replace
sensor.

Advertising