Operating Instructions
AX300-I
Teledyne
Analytical
Instruments
12
optional RS 232 Interface Cable (P/N B-75555) available from Teledyne for
connection to a standard RS 232 port on a computer or other suitably equipped
digital device.
CAUTION:
AX300-I ANALYZER SHOULD ONLY BE CONNECTED TO AN
EN60601-1/IEC60606-1 APPROVED DEVICE.
Figure 2-8: 0-1 VDC or RS 232 Digital Output Port
RS232 signal output has been tested with Windows HyperTerminal. The setup
changes with different windows versions. If other software other than windows
HyperTerminal is used, consult with factory.
Windows 95 version
Baud: 2400
Data:
7
Parity: ODD
Stop bit:
1
Flow control:
NONE
Windows 98, Windows 2000 and Window XP version:
Baud: 2400
Data: 8
Parity: ODD
Stop bit:
1
Flow control:
NONE