Teledyne AX300-I - Medical application oxygen analyzer User Manual

Page 24

Advertising
background image

Operating Instructions

AX300-I

Teledyne

Analytical

Instruments

16

troubleshooting section of this manual may provide additional assistance in lo-
cating the problem.

2.3.5 Anesthetic Gases

2.3.5.1 Gases That MAY INDUCE Reading Error

When using the R17MED sensor in the presence of anesthetic gases, the

oxygen reading may fall (see Table below). The magnitude of this error will de-
pend upon the level of oxygen and the duration of exposure.

The anesthetic agents listed in the following table (Halothane, Enflurane,

Isoflurane, Sevoflurane, and Desflurane) were vaporized into a stream of 30%
oxygen / 70% nitrous oxide, and the resulting drops in oxygen level after an ex-
posure of approximately two hours were noted.

Exposures in excess of two hours may produce slightly greater errors. The er-

rors listed are typical for all oxygen sensors such as the R17MED. Exposing the
sensor to air or gases that do not contain anesthetic agents for a period of time
equal to or greater than the exposure interval will eliminate the reading error in
most cases.

Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas

Gas or Vapor Level

(Balance: Mixture of 30% O2 / 70% N2O, except where noted)

Gas or Vapor

Test Level

Oxygen Reading

Error

Helium

50%, balance O2

0%

Nitrous Oxide

80%, balance O2

0%

Carbon Dioxide

10%, balance O2

0%

Halothane

4%

< 1.5% O2 *

Enflurane

5%

< 1.5% O2 *

Isoflurane

5%

< 1.5% O2 *

Sevoflurane

5%

< 1.5% O2 *

Desflurane

15%

< 1.5% O2 *


* Errors are approximate and may vary based on exposure times and concentra-

tions.
These performances meet or exceed the requirements of ISO 7767: 1997 (E).

Advertising