Skc real time monitor window temperatures display, Pump controls buttons – SKC DataTrac for the Pocket Pump User Manual

Page 11

Advertising
background image

9

SKC Real Time Monitor Window

Temperatures Display

The Temperatures display (Figure 7) shows the temperature data of the air entering the connected Pocket Pump.

Temperatures Display

Cell Readout

Min ........................................minimum air temperature during the program run
Max ........................................maximum air temperature during the program run
TWA .......................................time-weighted average of all air temperatures
Ambient ................................current air temperature

The Min, Max, and TWA are calculated from the temperatures measured during the total run time of the pump. Unless reset,
the temperature data will remain in memory and will be included in other Min, Max, and TWA calculations.

Reset Temps
To reset the minimum and maximum temperatures to ambient temperature and to reset the TWA, click once on the
Reset Temps butt on.

Before clicking the Reset Temps or the Reset Volume Temp and Time button, place the pump in Hold, otherwise
the TWA temperature may be erratic for the fi rst minute.

Pump Controls Buttons

The Pump Controls butt ons (Figure 8) directly control the connected pump.

Control Function
Run ............................places the pump in Run

Hold ..........................places the pump in Hold

Set Flow ....................opens the Set Flow window (Figure 16). Only

selectable when in Constant Flow mode; the
current fl ow rate is displayed beneath the butt on

Set CP ........................opens the Constant Pressure window (Figure 17).

Only selectable when in Constant Pressure mode; the current back pressure is displayed
beneath the butt on

Reset Volume
Temp and Time ........clears the accumulated volume, temperature, and run time

Fahrenheit ................selects the Fahrenheit temperature scale

Celsius .......................selects the Celsius temperature scale

in H

2

O .......................selects the inches H

2

O pressure units

mm Hg ......................selects the millimeters Hg pressure units

Standard ...................selects the standard display mode on the pump LCD (fl ow rate, time, and volume)

Enhanced ..................selects the enhanced display mode on the pump LCD (fl ow rate, time, volume, temperature,

and pressure)

C. Flow ......................selects Constant Flow mode; switching from Constant Pressure to Constant Flow mode

resets the Real Time Monitor display (Figure 5)

C. Press .....................selects Constant Pressure mode; switching from Constant Flow to Constant Pressure mode

resets the Real Time Monitor display (Figure 5)

Figure 7. Temperatures Display

Figure 8. Pump Controls Buttons