Getting Started

4) Flow Adjust Screw (No. 8)
Use the small bladed screwdriver attachment of the supplied toolkit to turn the flow adjust screw. To increase
the flow rate turn clockwise. The span of the flow adjust screw is 12 turns. DO NOT FORCE the flow adjust
screw, and ensure the screwdriver end is securely located in the adjust screw slot.
Use a suitable flow calibrator (such as a rotameter or Defender primary calibrator) to set the required flow rate,
do not use the built in flow indicator.

APPLICABLE TO MODELS 224-PCMTX4 & 224-PCMTX8 ONLY

5) START/HOLD Key (No. 3)
Once the pump has been switched on using the on/off switch, press the START/HOLD key to put the pump
into hold mode. The pump will stop running and the timer display will freeze. The legend HOLD is indicated
in the bottom left of the LCD screen. Press the START/HOLD key again to restart the pump and the timer will
continue to increase from the point at which it was frozen. The words SAMPLE RUNNING will appear in the top
right of the LCD screen. Subsequent pressing of this key will toggle between sample running and hold modes.

6) FLOW AND BATTERY CHECK Key (No. 2)
With the pump in hold mode, press the FLOW AND BATTERY CHECK key. The pump will start to run but the
timer will remain frozen at its current reading. The legend HOLD in the bottom left of the LCD screen will flash
on and off to indicate that the pump is in the flow and battery check mode. Flow checks can now be carried
out without the time taken being included on the timer display.
If the pump battery is fully charged the legend BATT OK is displayed. If the battery is flat the legend LO BATT
is displayed.
Press the FLOW AND BATTERY CHECK again to return the pump to hold mode.