Intrinsic Safety Information

The Universal MTX pump range

• Complies with the EC ATEX Directive 94/9/EC on equipment intended for use in potentially explosive

atmospheres, as intrinsically safe to EEx ia

IIC T4, for use in Europe.

• The equipment may be used in zones 1 and 2 with flammable gases and vapours with apparatus groups

IIA, IIB & IIC and with temperature classes T1,T2,T3,T4.

• The equipment is only certified for use in ambient temperatures in the range -20 to +40°C and should not

be used outside this range.

• The battery pack should not be charged in a hazardous area.
• Use only SKC approved chargers designated for the model.
• The equipment has not been assessed as a safety related device (as referred to in Directive 94/9/EC

Annex II, clause 1.5).

• The equipment should not be used if damaged in a way that could invalidate intrinsic safety. Such

defects might include cracking of the battery pack enclosure and internal encapsulant such that internal
components or cells are exposed.