Operating the Timer (224-PCMTX8 Only)

The timer function features three settings used to achieve the various modes of operation:
1. DELAYED START - The time period in minutes that will elapse prior to the pump starting to run. Setting

this value to zero disables the delayed start function.

2. SAMPLE PERIOD - The time period over which sampling will occur. Works in conjunction with the PUMP

PERIOD. Setting this value to zero will disable the timed run function and the pump will run continuously
until stopped using the START/HOLD key.

3. PUMP PERIOD - The time period in minutes during which the pump is actually running. If the PUMP

PERIOD is set the same or to a higher value than the SAMPLE PERIOD, the pump will run continuously
for the duration of the SAMPLE PERIOD. If the PUMP PERIOD is set to a lower value than the SAMPLE
PERIOD then the pump will automatically calculate on and off periods to apportion the required PUMP
PERIOD time over whole of the SAMPLE PERIOD time.

Setting a Simple Timed Run

Example: 8 hour timed run with no delayed start.
1. Switch the pump on and press the START/HOLD key to enter hold mode.
2. Press the SET-UP key to enter setup mode.
3. Press the MODE key to cycle to the SAMPLE PERIOD setting option. The right hand digit will be flashing.
4. The timer is set in minute values. Calculate the required sample period in minutes - 8 hours = 480 minutes.

Thefore the right hand digit should be set to 0.

5. Press the DIGIT SELECT key to select the second digit from the right.
6. Press the DIGIT SET key to increment the digit to 8.