Troubleshooting

Possible Fault

Corrective Action

Battery pack will not charge

• Check the battery charger by trying it with a different battery pack.

Replace the battery charger if required.

• Check the battery pack by connecting it to a known good battery

charger. If the battery pack still will not charge replace the battery pack.

Pump will not operate

• Check for faulty battery pack by trying a known good battery pack.

Replace the battery pack if required.

Pump flow faults continuously
(224-PCMTX4 & 224-PCMTX8
models only)

• Filter media back pressure too high. Try a lower flow rate and/or a less

restrictive filter media if the sampling method being used allows this.

• More than one filter accidentally fitted into sampling head. Ensure that

only one filter is fitted into the sampling head.

• Filter backing paper fitted in addition to / in place of filter media. Ensure

that only the filter is fitted into the sampling head.

• Pump particulate filter is blocked (appears black). Replace the

particulate filter.

• Tubing blocked or crimped. Replace tubing.

Pump cannot acheive required
flow rate

• Battery pack voltage low. Fully charge the battery pack.
• Filter media back pressure too high. Try a lower flow rate and/or a less

restrictive filter media if the sampling method being used allows this.

• Pump is in the low flow mode. Disable low flow mode.
• Pump valves leaking. Replace valve plate assemblies.