Preliminary procedure, Functional check – Teledyne LeCroy CP031 User Manual

Page 15

Advertising
background image

Instruction Manual

923362-00 Rev A

9

Description

Minimum Requirements

Test Equipment Examples

Calibration Fixture, 100 Turn

Loop

100 Turn loop in series with 0.5Ω ±
0.1% resistor with sense terminals

Teledyne LeCroy CP031-CF02

Calibration Fixture

ProBus Extension Cable

Teledyne LeCroy PROBUS-CF01

Banana Plug Adapter

Female BNC to Dual Banana Plug

Pomona 1269

Patch Cables (4 required)

Male Banana to Male Banana, 12"

Pomona B-12-0 (black)

Pomona B-12-2 (red)

BNC Adapter

BNC Male to Dual Banana Jack

Pomona 1296

Preliminary Procedure

1. Connect the CP031 to the channel 1 input of the oscilloscope and completely close the probe

slider.

2. Turn the oscilloscope on and allow at least 30 minutes warmup time for the CP031 and test

equipment before performing the Verification Procedure.

3. Turn on the other test equipment and allow these to warm up for the time recommended by

the manufacturer.

4. While the instruments are reaching operating temperature, make a photocopy of the

Performance Verification Test Record (located in Appendix A), and fill in the necessary data.

Functional Check

The functional check will verify the basic operation of the probe functions.

It is recommended that the Functional Check be performed prior to the Performance Verification
Procedure.

1. Select Channel 1 and verify that the probe bandwidth is set to Full (BWL OFF).

2. Verify that the CP031 is sensed and the CP031 tab appears behind the channel setup dialog.

3. On the CP031 tab, degauss the probe by touching DEGAUSS button, then OK.

4. Verify that "Performing Degauss on CP031...." is displayed at the bottom of the screen and no

error message remains displayed.