Teledyne LeCroy CP031 User Manual

Page 19

Advertising
background image

Instruction Manual

923362-00 Rev A

13

5. Set both DMMs to measure AC Volt.

6. With the CP031 removed from any signal and the slider returned to the LOCKED position,

degauss the probe by touching DEGAUSS (on the CP031 dialog), then OK.

7. Open the CP031 slider and position the probe input around the 100 Turn loop. Close and LOCK

the slider.

8. Adjust the Function Generator voltage until the voltage measured at the 'Current Sense'

terminals (DMM #2) is 50 mV ± 0.05 mV. (This corresponds to 10 A at the probe head).

9. Adjust variable resistor VR202 on the PCB until the voltage measured on DMM #1 is as close to

2X the voltage measured on DMM #2.

10. Verify that the measured voltage is between 0.990 V and 1.01 V.

11. Disconnect the probe from the test setup.

12. Re-assemble the compensation box by sliding the circuit board back into the box and inserting

and fastening the two screws.