CP031 Current Probe

2

923362-00 Rev A

Never install or remove the probe on bare conductors which are energized. The transformer
core and shield are grounded but not insulated and may contact the conductor when the locking
lever is open.

To avoid short circuits and accidents that could result in injury or death, use the CP031 current
probe only with conductors carrying 300 V or less.

To prevent short circuits and electric shock when conductors being measured carry less than
300 V but more than the safe voltage level (SELV-E) make sure that the conductors to be
measured are insulated with material conforming to:

o Overvoltage Category I, basic insulation requirement for working voltage of 300 V

o Pollution degree 2

Operating Environment

The accessory is intended for indoor use and should be operated in a clean, dry environment. Before
using this product, ensure that its operating environment is maintained within these parameters:

Temperature: 0° to 40° C

Humidity: ≤ 80% relative humidity (non-condensing)

Altitude: up to 2000 m (6562 feet)