Teledyne LeCroy Summit T3-16 PCIe Multi-lane Protocol Analyzer User Manual User Manual

Page 22

Advertising
background image

Teledyne LeCroy

Setting Up the Summit T3‐16 Analyzer using an Ethernet Connection

10

 

 

Summit T3‐16 PCI Express Multi‐Lane Protocol Analyzer User Manual

The PETracer software fills the list with devices that are connected over USB or 
discovered on the Ethernet network. The discovery mechanism works only within one 
network subnet. If a Summit is connected to the network on a different subnet, you can 
manually add the subnet to the list by clicking the Add Device button and specifying the 
IP address.

The Summit devices in the list are marked:

Locked: Some other client on the network is already connected to that device

Ready to connect: Available for connection

1. If a Summit device is marked Ready To Connect, you can select that device and press 

the Connect button to execute the connection procedure. 
After the connection is established, the application displays the Connection 
Properties dialog.

Figure 3.4: Connection Properties Dialog

2. Select an option:

Automatically connect to the device: When the application is started or when 
the named device is added to the network while the PETracer application is run‐
ning on this computer, the PETracer application will try to connect to the named 
device. 

Ask if I want to connect to the device: When the application is started or when 
the named device is added to the network while the PETracer application is run‐
ning on this computer, the PETracer application will display a message box allow‐
ing you to connect to the named device. 

Take no action: When you start the application or when you want to add the 
named device to the network while the PETracer application is running on this 
computer, you must connect manually to use the named device. 

Note:

When you close the application on this computer (or you perform a manual disconnect), the 

application disconnects from the device. 

3. Press OK in the Connection Properties dialog.

After you finish the connect procedure, the Summit device to which you have 
connected is marked as Ready and you can use it for recording (see 

Figure 3.5 on 

page 11

).

Advertising