13- Measure system frequency:


Use the "Set Address" routine to set address

DUT 1 = 0x7b

DUT2 = 0x7f

use write routine to write

0x1


Wait 15 ms

Use the "Set Address" routine to set address

DUT 1 = 0x78

DUT2 = 0xf8

use read routine to read

D[7..0]

Use the "Set Address" routine to set address

DUT 1 = 0x79

DUT2 = 0xf9

use read routine to read

D[15..8]

Use the "Set Address" routine to set address

DUT 1 = 0x7a

DUT2 = 0xfa

use read routine to read

D[23..16]


system frequency = (D[23..16] D[15..8] D[7..0])*100/1048575 MHz

For example if the read numbers were 0x5, 0x 55 & 0x46 then 0x55546 = decimal 349510 *100/1048575 = 33.33 MHz.

14- Measure DUT Voltages and current:


Select the voltage or current to convert
Use the "Set Address" routine to set address

DUT 1 = 0x67

DUT2 = 0xe7

use write routine to write

D[7..0] Signal to Convert & Read


0X08

+2.5V Reference

0X09

+12V UUT

0X0A

+5V UUT

0X0B

+3.3V UUT

0X0C

+5V Aux

0X0D

Current +5V

0X0E

Current +12V

0X0F

Current +3.3V

Wait 10 ms after

selecting the signal



Start Conversion

Use the "Set Address" routine to set address

DUT 1 = 0x68

DUT2 = 0xe8

use extended write routine to write.

0x0


Wait 2 ms

To read data, shift data to a parallel register
Use the "Set Address" routine to set address

DUT 1 = 0x69

DUT2 = 0xe9

use write routine.

0x0


Repeat this write 13 times.

To read converted Data out
Use the "Set Addres s" routine to set address

DUT 1 = 0x6a

DUT2 = 0xea

use read routine to read the data.

Use the "Set Address" routine to set address

DUT 1 = 0x6b

DUT2 = 0xeb

use read routine to read the data.