S-parameter files path, Stop on test to review results, Test mode – Teledyne LeCroy QPHY-USB3-Tx-Rx User Manual

Page 39: Use pert3, Use rf switch, Pert3 setup, Pert3 hostname or ip address, Pert3 ssc enabled, Transmitter tests, Deskew measure mode

Advertising
background image

QPHY-USB3-Tx-Rx Software Option

917719 Rev

A

39

S-Parameter Files Path

This variable allows the user to specify the path on the oscilloscope that contains the S-Parameter files used for
emulation and/or de-embedding. The default for this variable is

“D:\Applications\EyeDr\USB3. ”

SSC
This variable allows the user to specify whether or not Spread Spectrum Clocking (SSC) is enabled on the
Product Under Test. The SigTest Average UI test has different limits based on whether or not SSC is enabled. By
setting this variable, the correct limit for this test can be used. Select

“Enabled” if Spread Spectrum Clocking is

enabled, otherwise select

“Disabled.” The default is “Enabled” because SSC is required by the USB 3.0

Specification.

Stop On Test to review results

This variable gives the user the option to enable Stop on Test mode. When set to

“Yes”, the testing pauses after

each test allowing the us er to review the results. The setup is saved so oscilloscope settings can be modified by
the user. When execution is resumed, the setup is recalled. Note that any new acquisition completed while the
testing is paused may cause the testing to misbehave. The default value for this variable is

“No.”

Test Mode

This variable allows the user to decide whether the tests should be run on newly acquired data or if previously
saved waveforms shoul d be us ed. The choices for this variable are

“Acquire New Data” and “Use Saved Data.” If

“Use Saved Data” is selected, the waveforms are recalled from the location specified in the Saved Waveform
Path
variable. The default value for this variable is

“Acquire New Data.”

Use PeRT3

This variable allows the users to select if they would like QualiPHY to automate the PeRT

3

to configure the

Product under test for each test. If this variable is set to

“No”, the user must have the ability to stimulate their

Product under test to transmit the required test patterns for transmitter testing. The default setting for this variable
is

“No.”

Use RF Switch

This variable allows the users to specify whether or not they have an RF Switch. If this variable is set to

“Yes,” the

oscilloscope communicates with the RF Switch and makes the required connections without any user interaction.
If this variable is set to

“No,” the user must connect their signals when prompted. The default value for this

variable is

“No.”

PeRT3 Setup

PeRT3 Hostname or IP Address
This variable allows the user to enter the hostname or IP Address of the PeRT3 Server (the machine that the
PeRT

3

is connected to via USB). If QualiP HY is running from a remot e PC and not from the oscilloscope,

“localhost refers to the oscilloscope. The default value for this variable is “localhost.”

PeRT3 SSC Enabled

This variable allows the user to enable SSC on the PeRT

3

’s clock output. This is required for receiver testing. The

default for this variable is

“Yes.”

Transmitter Tests

Deskew measure mode
This variable allows the user to select the method for deskewing their cables . The choic es for this variable are
“Wizard Method” and “User Defined.” “Wizard Measured” is used to launch the automated deskew measurement.
“User Defined” is used if the us er would like to set the Deskew value in picoseconds variable when the deskew
value is already known. The default value for this variable is

“Wizard Method.”