MANUAL CALIBRATION PROCEDURES

Cable Deskewing using the Fast Edge Output (WavePro 7 Zi and WaveMaster 8Zi only)

The following procedure demonstrates how to manually deskew two oscilloscope channels and cables using the
fast edge out put, with no need for any T connector or adapters.

This can be done once the temperature of the oscilloscope is stable. The oscilloscope must be warmed up for at
least a half-hour before proceeding. This proc edure sho uld be run again if the temperature of the oscilloscope
changes by more than a few degrees.

For the purpose of this procedure, the two channels being deskewed are referred to as Channel X and Channel
Y. The reference channel is Channel X and the channel being deskewed is Channel Y.

1. Begin by rec alling the Default Oscilloscope Setup

2. Configure the oscilloscope as follows:

Timebase

i. Fixed Sample Rate

ii. Set the Sample Rate to 40 GS/s

iii. Set the Time/Division to 1 ns/div

Figure 32 - Timebase Settings for Deskew with the Fast Edge Output

Channels

i. Turn on Channel X and Channel Y.

ii. Set V/div for Channel X and Channel Y to 100mV/div.

iii. Set the A veraging of Channel X and Channel Y to 500 sweeps .

iv. Set the Interpolation of Channel X and Channel Y to Sinx/x.

Figure 33 - Channel Pre-Processing Settings for Deskew with the Fast Edge Output

Trigger

i. Configure to Source to be Fa stEdge.

ii. Set the Slope to Positive.