4 process current, 5 process voltage – NEWPORT iDRA Controller User Manual

Page 14

Advertising
background image

10

2.3.4 Process Current

The figure below shows the wiring hookup for Process Current 0 – 20 mA.

Figure 2.6 Process Current Wiring Hookup

(Internal and External Excitation)

When configuring your instrument, select Process Type in the Input Type Menu
(see Part 3).

2.3.5 Process Voltage

The figure below shows the wiring hookup for Process Voltage 0 – 100 mV,
0 – 1 V, 0 – 10 V.

Figure 2.7 Process Voltage Wiring Hookup

(with Sensor Excitation and without Sensor Excitation)

R

L

- Voltage limited resistor, which allows to convert 24 Vdc internal excitation

voltage to the appropriate process input value. For instance: if the potentiometer
value is equal to 10 kΩ, the minimum R

L

is 14 kΩ for 10 V process input.

When configuring your instrument, select Process Type in the Input Type Menu
(see Part 3).

100 mV

Return

1 V or 10 V

R

L

100 mV

1 V or 10 V

Return

0-20 mA

Internal

Excitation

0-20 mA

External

Excitation