Fully Automatic AED Plus Administrator’s Guide

21

• displays a red “X” in the status indicator window, indicating that the batteries are at low capacity

or that the Fully Automatic AED Plus has failed other self tests.

Table 3: Battery Condition

Battery Condition

Indications

Correction

Low Battery with Fully

Automatic AED Plus off.

Audible beep from Fully

Automatic AED Plus once every

minute.

Replace batteries.

Low Battery during power

up self test.

CHANGE BATTERIES prompt

(when Fully Automatic AED

Plus is powered on)

Replace batteries.

Low Battery or other self

test failure with Fully

Automatic AED Plus

powered off or during self

test.

Status Indicator has red “X”

indicating failure to operate

(when off).

Replace batteries. Check or

replace electrodes. If red “X”

remains, return to ZOLL

Technical Support for service.

Low Battery with Fully

Automatic AED Plus

powered on.

CHANGE BATTERIES prompt

(unit powered on).

Replace batteries as soon as

possible.

Dead Battery

Status Indicator has red “X”

indicating failure to operate

(when off).

Replace batteries. If red “X”

remains, return to ZOLL

Technical Support for service.