ZOLL AED Plus Fully Automatic Rev B User Manual

Page 42

Advertising
background image

32

Fully Automatic AED Plus Administrator’s Guide

Immunity test

IEC 60601 test

level

Compliance

level

Electromagnetic environment -

guidance

Voltage dips,

short

interruptions and

voltage variations

on power supply

input lines.

IEC 61000-4-11

<5% Ut (>95% dip

in Ut) for 0.5 cycle

40% Ut (60% dip in

Ut) for 5 cycles

70% Ut (30% dip in

Ut) for 25 cycles

<5% Ut (>95%dip

in Ut) for 5 sec

Not applicable

Not applicable

Not applicable

Not applicable

Power frequency

(50/60 Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields

should be at levels characteristic of a

typical location in a typical

commercial or hospital environment

NOTE Ut is the a.c. mains voltage prior to application of the test level.

The Fully Automatic AED Plus is intended for use in the electromagnetic environment specified
below. The customer or the user of the Fully Automatic AED Plus should assure that it is used
in such an environment.

Immunity test

IEC 60601 test

level

Compliance

level

Electromagnetic environment -

guidance

Portable and mobile RF

communications equipment should be

used no closer to any part of the Fully

Automatic AED Plus, including

cables, than the recommended

separation distance calculated from

the equation applicable to the

frequency of the transmitter.

Recommended separation distance

Conducted RF

IEC 61000-4-6

3 Vrms

150 kHz to 80 MHz

outside ISM bands

a

3 Vrms

d = 1.17

10 Vrms

150 kHz to 80 MHz

in ISM bands

a

10 Vrms

d = 1.20

P

P