Fully Automatic AED Plus Administrator’s Guide

Troubleshooting

The following table summarizes common error indications on the Fully Automatic AED Plus, and their

associated corrective action. Return the Fully Automatic AED Plus to ZOLL’s Technical Service

Department if the Fully Automatic AED Plus is not working properly.

Table 5: Troubleshooting

Technical Problem

Recommended Action

Self test failed.

Initiate manual self test by pressing and holding the

ON/OFF Button for more than 5 seconds. Attempt

to repair the device by replacing the batteries or

electrodes. If the Fully Automatic AED Plus fails

test again, remove the Fully Automatic AED Plus

from service and contact ZOLL Technical Service.

CHANGE BATTERIES prompt.

Replace all batteries with new batteries at the same

time. Press the Battery Reset Button when

prompted.

Red “X” in status indicator window.

Initiate manual self test by pressing and holding the

ON/OFF Button for more than 5 seconds.

Check to see if cable is attached properly to the

Fully Automatic AED Plus or replace the

electrodes.

Cycle power on the Fully Automatic AED Plus by

turning the unit off, then on again.

Replace all batteries at the same time with new

batteries that are less than 1 year old. Press the

Battery Reset Button when prompted.

If the Fully Automatic AED Plus still does not

operate correctly, remove unit from service and

contact ZOLL Technical Service.

Beeping noise when Fully Automatic

AED Plus is off.

Remove Fully Automatic AED Plus from service

and replace batteries. Replace all batteries at the

same time with new batteries. Press the Battery

Reset Button when prompted.

If beeping continues, contact ZOLL Technical

Service.

PLUG IN CABLE prompt.

Check cable connection between electrodes and the

Fully Automatic AED Plus.

ANALYSIS HALTED. KEEP VICTIM

STILL prompt.

Excessive artifact detected during ECG analysis.

Victim must be motionless during ECG analysis. Do

not touch the victim during analysis. Keep the

victim still. If the rescuer is using the Fully

Automatic AED Plus in an emergency vehicle,

bring the vehicle to a halt before performing ECG

analysis.