ZOLL AED Plus Fully Automatic Rev B User Manual

Page 41

Advertising
background image

Fully Automatic AED Plus Administrator’s Guide

31

Guidance and Manufacturer’s Declaration - Electromagnetic
Emissions

Table 8: EMC Specifications

The Fully Automatic AED Plus is intended for use in the electromagnetic environment specified
below. The customer or the user of the Fully Automatic AED Plus should assure that it is used in
such an environment.

Emissions test

Compliance

Electromagnetic environment -

guidance

RF emissions CISPR11

Group 1

The Fully Automatic AED Plus uses

RF energy only for its internal

function. Therefore, its RF emissions

are very low and are not likely to

cause any interference in nearby

electronic equipment.

RF emissions CISPR 11

Class B

Harmonic Emission

IEC 61000 3-2

Not applicable

Voltage Fluctuations/Flicker

Emission

IEC 61000 3-3

Not applicable

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and

put into service according to EMC information provided in this document.

The Fully Automatic AED Plus is intended for use in the electromagnetic environment specified
below. The customer or the user of the Fully Automatic AED Plus should assure that it is used
in such an environment.

Immunity test

IEC 60601 test

level

Compliance

level

Electromagnetic environment -

guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or

ceramic tile. If floors are covered with

synthetic material, the relative

humidity should be at least 30%.

Electrical fast

transient/burst

IEC 61000-4-4

± 2 kV for power

supply lines

± 1 kV for input/

output lines

Not applicable

± 1 kV I/O

Surge

IEC 61000-4-5

± 1 kV differential

mode

+/- 2 kV common

mode

Not applicable

Not applicable