Fully Automatic AED Plus Administrator’s Guide

Maintenance and Troubleshooting

This section describes the following functions to maintain the Fully Automatic AED Plus:

• Maintaining the Fully Automatic AED Plus

• Cleaning the Fully Automatic AED Plus

• Optional Maintenance for Technical Professionals

• Troubleshooting

Maintaining the Fully Automatic AED Plus

• Inspect frequently, as necessary.

• Check for the green check (

)

showing that the Fully Automatic AED Plus is ready to use.

• Verify that electrodes are within their expiration date.

• Verify that batteries are within their expiration date.

• Verify that electrodes are pre-connected to the input connector.

• Verify that supplies are available for use (razor, mask, gloves, extra batteries.)

Maintenance Checklist

Use the following maintenance checklist when you periodically check your Fully Automatic AED Plus.

Table 4: Maintenance Checklist

Cleaning the Fully Automatic AED Plus

• After each use, clean and disinfect the Fully Automatic AED Plus with a soft, damp cloth using

90% isopropyl alcohol, or soap and water, or chlorine bleach and water mixture (30 ml/liter water).

• Do not immerse any part of the Fully Automatic AED Plus in water.

• Do not use ketones (MEK, acetone, etc.) to clean the Fully Automatic AED Plus.

• Avoid using abrasives (e.g., paper towel) on the display window or IrDa port.

• Do not sterilize the Fully Automatic AED Plus.

Check the following

Pass

Fail

Is the unit clean, undamaged, free of excessive wear?

Are there any cracks or loose parts in the housing?

Verify electrodes are connected to the Fully Automatic AED Plus and

sealed in their package. Replace if expired.

Are all cables free of cracks, cuts and exposed or broken wires?

Turn the Fully Automatic AED Plus on and off and verify the green check

indicates ready for use.

Batteries within expiration date. Replace if expired.

Check for adequate supplies.



