8.0 Calibrating Pulses on Strip Chart Test

Tools Needed:

None

Test Setup:

None

.

Do this...

Verify that...

Pass/Fail/
NA

8.1

Press the RECORDER button.

8.2

Press and hold SIZE button to
activate the calibration signal.

The strip chart displays a signal of 300 ppm with an
amplitude of 10 mm +/- 1 mm. The signal also appears
on the video display.

o

o

o

Record your results on the Maintenance Tests Checklist.