0 calibrating pulses on strip chart test, M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 29
Advertising
M Series Service Manual
19
8.0 Calibrating Pulses on Strip Chart Test
Tools Needed:
None
Test Setup:
None
.
Do this...
Verify that...
Pass/Fail/
NA
8.1
Press the RECORDER button.
8.2
Press and hold SIZE button to
activate the calibration signal.
The strip chart displays a signal of 300 ppm with an
amplitude of 10 mm +/- 1 mm. The signal also appears
on the video display.
o
o
o
Record your results on the Maintenance Tests Checklist.