M Series Service Manual

36

18.0 EtCO

2

Monitor Test (for EtCO

2

Option)

Tools Needed:

Novametrix Capnostat Simulator Tb 1265/7100.

Test Setup:

Install the battery

On the Novatrix Simulator, set the folling

Set inspired CO

2

to OFF

Set% CO

2

to 0

Set Sensor Location to ZERO CELL

Set Source Current to NORMAL

Set CO

2

mode to CONTINUOUS. Set Temperature to NORMAL

Attach simulator to the unit.

Do this...

Verify that...

Pass/
Fail

18.1

Turn the selector switch to
MONITOR mode (For AED units,
turn the selector switch to ON,
and select Manual mode.) Attach
the EtCO

2

simulator to the M

Series input connection.

CO

2

SENSO WARMUP message displays.

Note: You may need to wait up to 5 minutes for the
warm-up message to disappear.

o

o

18.2

Perfom manual zeroing. Press
PARAM softkey, select EtCO

2

,

press ENTER, and then press
ZERO softkey, then Enter.

Verify “ZEROING CO2” Adapter” is displayed, then
changes to “ZERO DONE” when complete.

o

o

18.3

On the Novametrix SImulator, set
SENSOR LOCATION to AA
CELL, set% CO

2

to 10% CO

2

mode to RESPIRATION.

Verify the EtCO

2

reading of 74-84mmHg appears on the

M Series display.

Note that you may need to wait up to 10 seconds for the
unit to stabilize.

o

o