0 bluetooth test, Bluetooth test, M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 61

Advertising
background image

M Series Service Manual

51

24.0 Bluetooth Test

Tools Needed:

USB Bluetooth dongle with driver CD, Bluetooth application software BlueSoleil.

Test Setup:

Verify that the M Series unit is configured as follows: Bluetooth INSTALLED, and Baud Rate set to 115200 or 38400.

Do this...

Verify that...

Pass/
Fail

24.1

Power up the M Series unit while
pressing in and holding the left-
most softkey.

The M Series unit displays the System Utilities Menu.

o

o

24.2

Insert a PCMCIA card containing
data into the card reader.

24.3

Press the UPLOAD CARD
softkey, then press the softkey
UPLOAD CARD
.

Verify that the Bluetooth LED is on and displays a green
light.

o

o

24.4

Press the SEND softkey.

24.5

Double click on the HyperTerminal
shortcut (BT115200) from the
desktop.

Note:

If setting up Hyperterminal
for the first time set COM
PORT programmed for
Bluetooth adapter,
115200bps, 8, N, 1, N

24.6

Start Bluetooth Utility software
(Blue Soleil) on the computer.

24.7

In the IVT Corporation Blue Soleil
window under “View”, select
refresh devices.