0 nibp monitor test, Nibp monitor test, M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 57

Advertising
background image

M Series Service Manual

47

22.0 NIBP Monitor Test

The NIBP monitor test verifies the repeatability of the systolic, diastolic, and mean blood pressure measurements, as
well as the patient pulse rate calculation.

Tools Needed:

Fluke NIBP simulator (the values and procedure that we provide are specific to the BP Pump 2).

Note: The primary propose of an NIBP simulator is to reproduce a pressure profile similar to a live patient to be
used for testing t h e repeatability and functionality of the system. There are many different NIBP simulators on
the market, each manufacturer uses a different method to develop their algorithm. Consequently, readings from
different simulators may vary. To test for repeatability, you should first establish the offset

1

of your simulator. The

offset value should then be used to determine the expected values. NIBP simulators cannot be used as a source for
testing the accuracy of the non-invasive blood pressure measurements of devices such as the ZOLL M Series
monitor/defibrillator.

Test Setup:

Connect the Analyzer’s simulator hose to the NIBP connector on the M Series unit.

Set the following parameters on the NIBP Analyzer:

2

Make sure the ECG cable is not connected to the M Series unit.

If the SpO2 option is installed, make sure that the Masimo cable is NOT connected to the M Series unit.

1

NIBP Simulators may produce a reading on the NIBP monitor that is shifted from the simulator's setting. The offset value must be

established based on a statistical sample of monitors and readings. Please contact ZOLL Technical Support if you require assis-
tance establishing the offset of the simulator and test set-up that you are utilizing.

2

If you are using the Fluke® BiomedicalCuftlink, you must change the shift value of the Blood Pressure Envelope to +3 on the Pres-

sure Curve Adjust Menu.

3

Not all simulators have a setting of 93mmHg. Check the simulators user's manual for recommendations.

Table 1-1:

Parameter

Value

Systolic pressure

120 mmHg

Diastolic pressure

80 mmHg

Mean pressure

93 mmHg

3

Heart pressure

80 bpm