0 summary report test, M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 39

Advertising
background image

M Series Service Manual

29

14.0 Summary Report Test

Tools Needed:

QED 6 Defibrillator Analyzer.

Test Setup:

Connect the universal cable to the QED 6 analyzer.

If you are using paddles, place the paddles on the analyzer’s discharge plates.

Do this...

Verify that...

Pass/
Fail

14.1

Press and hold the SUMMARY
softkey for 4 to 8 seconds to
display erase options. Press the
ERASE SUMMARY softkey to
erase any previously stored data
erase any previously stored data.

The unit displays the message, ERASING REPORT.

o

o

14.2

Set selector switch to DEFIB.
Select 200J (360J for
Monophasic) using the ENERGY
SELECT
button, and press the
CHARGE
button. When charged,
press the SHOCK button to
discharge into the defibrillator
analyzer.

The unit successfully discharges and prints a strip chart.

o

o

14.3

Wait 18 seconds, then press the
Code Marker softkey. Press the
CPR softkey.

The Code Markers appear.

o

o

14.4

Turn the unit off. Wait 10 seconds
and then turn the unit on. Press
the SUMMARY softkey, then
press the PRINT CHART softkey,
then press PRINT ALL.

The summary report prints. The report displays the
correct date, time, 1 shock delivered and Code Marker
event.

o

o

Record your results on the Maintenance Tests Checklist.