System messages, Selecting the heart rate/ pulse rate (hr/pr) tone – ZOLL Propaq MD Rev A User Manual
Page 121
SpO
2
System Messages
9650-0805-01 Rev. A
Propaq MD Operator’s Guide
9-9
Selecting the Heart Rate/ Pulse Rate (HR/PR) Tone
The unit allows you enable or disable the tone that the monitor uses to indicate detection of the
patient’s pulse:
On
or
Off
(no tone sounds). The default tone is
Off
.
SpO
2
System Messages
The Propaq MD unit may display the following system messages when monitoring SpO
2
:
System Message
Cause
INITIALIZING
The SpO
2
pulse oximeter is initializing.
SEARCHING
The unit is searching for a pulse.
CHECK SENSOR
The SpO
2
sensor has become disconnected from the
unit, or the sensor is no longer on the patient. Check
sensor and then reconnect it to the unit or reapply it to the
patient.
SENSOR FAILURE
The SpO
2
sensor is defective. Replace the sensor.
LOW PERFUSION
Perfusion has dropped below 20%.
SPO2 DISABLED
A system error has occurred. The Propaq MD unit cannot
take SpO
2
measurements and should be serviced.