System messages, Selecting the heart rate/ pulse rate (hr/pr) tone – ZOLL Propaq MD Rev A User Manual

Page 121

Advertising
background image

SpO

2

System Messages

9650-0805-01 Rev. A

Propaq MD Operator’s Guide

9-9

Selecting the Heart Rate/ Pulse Rate (HR/PR) Tone

The unit allows you enable or disable the tone that the monitor uses to indicate detection of the
patient’s pulse:

On

or

Off

(no tone sounds). The default tone is

Off

.

SpO

2

System Messages

The Propaq MD unit may display the following system messages when monitoring SpO

2

:

System Message

Cause

INITIALIZING

The SpO

2

pulse oximeter is initializing.

SEARCHING

The unit is searching for a pulse.

CHECK SENSOR

The SpO

2

sensor has become disconnected from the

unit, or the sensor is no longer on the patient. Check
sensor and then reconnect it to the unit or reapply it to the
patient.

SENSOR FAILURE

The SpO

2

sensor is defective. Replace the sensor.

LOW PERFUSION

Perfusion has dropped below 20%.

SPO2 DISABLED

A system error has occurred. The Propaq MD unit cannot
take SpO

2

measurements and should be serviced.