Recorder, Battery – ZOLL Propaq MD Rev A User Manual

Page 221

Advertising
background image

Recorder

9650-0805-01 Rev. A

Propaq MD Operator’s Guide

A-17

Recorder

Type: High-resolution thermal array.

Annotation: Time, date, ECG lead, ECG gain, heart rate, defibrillation and pacing parameters
and treatment summary events.

Paper Width: 50 mm.

Paper Speed: 25 mm/sec, 50 mm/sec

Delay: 6 seconds.

Frequency Response: Automatically set to monitor's frequency response.

Treatment Summary:

10 switches to record key ACLS events (IV, INTUB, EPI, LIDO, ATROP, etc.).
Automatically logs into memory the type of event, time and ECG sample.

Patient Data Log: A combination of 24 hours of trends at 1-minute intervals, 1000 time-
stamped events, and 32 snapshots, including monitor, defibrillator, pacer, and treatment
snapshots. The actual number of trends, events, or snapshots stored could be more or less
depending on the use profile.

Record Modes: Manual and automatic (User-configurable).

Battery

Type: Rechargeable Lithium-Ion, 11.1Vdc, 6.6 Ah, 73Wh

Capacity:

With a new, fully charged battery operating at room temperature:

At least

6 hours of continuous monitoring of ECG, SpO2, CO2, three Invasive Pressure

channels, and 2 channels of Temperature, with NIBP measurements every 15 minutes and
10 200 J shocks (display set to 30%).

At least 100 discharges at maximum shock energy (200 joules).

At least 3.5 hours pacing, with ECG, SPO2, CO2, three Invasive Pressures, temperature,
NIBP every 15 minutes and pacing at 180 ppm, and 140 mA.

At least 10 discharges at maximum shock setting (200 joules) after a Low Battery
indication.

Note:

Proper battery care is required to maintain maximum available capacity.

Battery Indicators:

5 Battery capacity LED indicators, Fault indicator, Recalibration indicator

Recharge Rate: 100% in 4 hours, when initiated at Low Battery indication.

Advertising