ZOLL Propaq MD Rev A User Manual

Page 233

Advertising
background image

Electromagnetic Compatibility Guidance and Manufacturer’s Declaration

9650-0805-01 Rev. A

Propaq MD Operator’s Guide

A-29

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity (IEC
60601-1-2 Table 203)

The Propaq MD unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the Propaq MD should assure that it is used in such an
environment.

Immunity Test

IEC 60601 Test Level

Compliance
Level

Electromagnetic Environment – Guidance
IEC 60601-1-2

Portable and mobile RF communications
equipment should be used no closer to any
part of the Propaq MD unit, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.

Recommended Separation Distance

Conducted RF
IEC 61000-4-6

3 Vrms
150 kHz to 80 MHz
outside ISM bands

a

3 Vrms

d = 1.17

10 Vrms
150 kHz to 80 MHz in
ISM bands

a

10Vrms

d = 1.2

Radiated RF
IEC 61000-4-3

10 V/m
80 MHz to 2.5 GHz

10 V/m

d = 1.20
80 MHz to 800 MHz

d = 2.30
800 MHz to 2.5 GHz

3 V/m
80 MHz to 2.5 CHz
(IEC 60601-2-34)

3 V/m
(IBP only)

d = 4
80 MHz to 800 MHz

d = 7.67
800 MHz to 2.5 GHz

Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance in meters (m).

b

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey

c

, should be

less than the compliance level in each frequency range

d

.

Interference may occur in the vicinity of equipment marked with the following symbol:

P

P

P

P

P

P