ZOLL Propaq MD Rev A User Manual

Page 231

Advertising
background image

Electromagnetic Compatibility Guidance and Manufacturer’s Declaration

9650-0805-01 Rev. A

Propaq MD Operator’s Guide

A-27

Electromagnetic Compatibility Guidance and
Manufacturer’s Declaration

In-Flight Use (RTCA/DO-160):

The Propaq MD unit complies with RTCA/DO-160, Environmental Conditions and Test
Procedures for Airborne Equipment, using the methods in Section 21, Category M for Radiated
and Conducted Radio Frequency Energy.

General Use Environments (IEC 60601-1-2):

The Propaq MD unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the Propaq MD unit should assure that it is used in such an
environment.

Emissions Test

Compliance

Electromagnetic Environment – Guidance

RF Emissions CISPR 11

Group 1

The Propaq MD unit uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.

RF Emissions CSPR 11

Class B

Harmonic Emission IEC
6100-3-2

Class A

The Propaq MD unit is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.

Voltage Fluctuations/ Flicker
Emissions IEC 61000-3-3

Complies

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put
into service according to EMC information provided in this document.

Advertising