Warning, 3) installation – Yokogawa EJX115A User Manual

Page 14

Advertising
background image

<2. Handling Cautions>

2-6

IM 01C25T04-01EN

Note 4. Specific Conditions of Use

WARNING

• Electrostatic charge may cause an explosion

hazard. Avoid any actions that cause the

gerenation of eletrostatic charge, such as

rubbing with a dry cloth on coating face of

the product.

• When the lightning protector option is

specified, the apparatus is not capable

of withstanding the 500V insulation test

required by EN60079-11. This must be taken

into account when installing the apparatus.

(2) Electrical Connection
A mark indicating the electrical connection type is

stamped near the electrical connection port. These

marks are as follows.

F0204.ai

Location of the mark

Screw Size

Marking

ISO M20 × 1.5 female

ANSI 1/2 NPT female

M

N or

W

A*

1

or

*1: Not applicable for option code /KF22.

(3) Installation

WARNING

• All wiring shall comply with local installation

requirements and the local electrical code.

• There is no need for a conduit seal in

Division 1 and Division 2 hazardous

locations because this product is sealed at

the factory.

(4) Operation

WARNING

• OPEN CIRCUIT BEFORE REMOVING

COVER. INSTALL IN ACCORDANCE WITH

THIS USER’S MANUAL

• Take care not to generate mechanical

sparking when accessing the instrument and

peripheral devices in a hazardous location.

(5) Maintenance and Repair

WARNING

The instrument modification or part replacement

by other than an authorized Representative of

Yokogawa Electric Corporation is prohibited and

will void the certification.

(6) Name Plate

Name plate

Tag plate for flameproof type

Tag plate for intrinsically safe Ex ia

Tag plate for intrinsically safe Ex ic

F0205.ai

AFTER DE-ENERGIZING, DELAY 5 MINUTES

BEFORE OPENING.

WHEN THE AMBIENT TEMP. ≥ 65°C,

USE THE HEAT-RESISTING CABLE AND CABLE GLAND ≥ 90°C

POTENTIAL ELECTROSTATIC CHARGING HAZARD

No. KEMA 07ATEX0109 X

Ex d IIC T6...T4 Gb, Ex tb IIIC T85°C Db

Enlcosure : IP66, IP67

TEMP. CLASS T6 T5 T4

MAX PROCESS TEMP.(Tp.) 85 100 120 °C

Tamb. -50 to 75 80 75 °C

T85°C(Tamb.:-30(-15) to 75°C, Tp.:85°C)(for Dust)

WARNING

Ex ic IIC T4 Gc

IP66

Tamb -30(-15) TO 60°C

MAX. PROCESS TEMP. 120°C

Ui=32V, Ci=3.52nF, Li=0µH

POTENTIAL ELECTROSTATIC CHARGING HAZARD - SEE USER'S MANUAL.

D

WARNING

No. KEMA 04ATEX1116 X

Ex ia IIC/IIB T4 Ga Ta:-40 to 60°C

Ex ia IIIC T85°C T100°C T120°C Db Ta:-30(-15) to 60°C

MAX PROCESS TEMP.(Tp.):120°C

T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)

Enclosure: IP66/IP67

FISCO field device(IIC/IIB)

Entity parameters Ui=24V, Ii=250mA, Pi=1.2W, Ci=3.52nF, Li=0μH

WARNING

POTENTIAL ELECTROSTATIC CHARGING HAZARD - SEE USER'S MANUAL.

D

*3

*3

Advertising