<8. Diagnostic Information>

8-4

IM 01C25T04-01EN

Table 8.3

Contents of DIAGNOSIS_EXTENSION

Octet Bit

Category

Display Message

Description

NAMUR

NE107

Category

Default

Mask

*1

1

0

System Alarm Pressure Sensor Failure (AL.01)

Pressure sensor problem

F

1

1

Capsule Temp Sensor Failure (AL.01)

Capsule temperature sensor problem

F

1

2

Capsule EEPROM Failure (AL.01)

Capsule memory problem

F

1

3

AMP Temp Sensor Failure (AL.02)

Amplifier temperature sensor problem

F

1

4

AMP EEPROM Failure (AL.02)

Amplifier memory problem

F

1

5

CPU Board Failure 3,4 (AL.02)

Amplifier memory problem

F

1

6

CPU Board Failure 5 (AL.02)

Amplifier memory problem

F

1

7

CPU Board Failure 1 (AL.02)

Amplifier memory problem

F

1

2

2

Process Alarm Diff Pressure Outside Range Limit (AL.10) Input Pressure is outside measurement

range limit of capsule

S

1

3

Static Pressure Outside Range Limit

(AL.11)

Static Pressure exceeds limit

S

1

4

Capsule Temp Outside Range Limit

(AL.12)

Capsule Temerature is out of range

S

1

5

AMP Temp Outside Range Limit (AL.13) Amplifier Temerature is out of range

S

1

3

0

Setting Alarm

Diff Pressure Span /Zero Trim Error

(AL.50)

Trimming range error for differential

pressure span

C

1

1

Static Pressure Span/Zero Trim Error

(AL.51)

Trimming range error for static pressure

span

C

1

2

STB Configuration Error (AL.60)

Setting of Sensor TB is abnormal

C

0

4

Warning

FB Simulation Active

(AL.72, 90, 91, 92, 93, 94, 95 )

Simulation function of AI1 block is active

Simulation function of AI2 block is active

Simulation function of AI3 block is active

AI1 in MAN mode

AI2 in MAN mode

AI3 in MAN mode

Totalizer block in MAN mode

C

1

5

STB Simulation Active (AL.96)

Simulation function of Sensor TB is active

C

1

4

1

Mode Alarm

Mode Check

(AL.40, 41, 43, 44, 45, 72)

PB in O/S mode

Sensor TB in O/S mode

AI1 in O/S mode

AI2 in O/S mode

AI3 in O/S mode

C

0

4

Process Alarm AI1 Upper/Lower Limit Alarm (AL.30)

AI1 Hi Hi Alarm occurs

AI1 Lo Lo Alarm occurs

---

1

5

AI2 Upper/Lower Limit Alarm (AL.31)

AI2 Hi Hi Alarm occurs

AI2 Lo Lo Alarm occurs

---

1

6

AI3 Upper/Lower Limit Alarm(AL.32)

AI3 Hi Hi Alarm occurs

AI3 Lo Lo Alarm occurs

---

1

7

Totalizer Upper/Lower Limit Alarm (AL.33) TOT Hi Hi Alarm occurs

TOT Lo Lo Alarm occurs

---

1

*1: Default value of DIAGNOSIS_MASK_EXTENSION_RW.