<6 OPERATION>

6-28

IM 01R01A11-01EN

6.3.2

Pulse Output

This item checks the pulse output from the converter.
Check the Suffi x Code of AXF converter or AXF integral fl owmeter.

■ Without optional code EM, see “(A) Pulse Output (without optional code EM)”.
■ With optional code EM, see “(B) Active Pulse Output (with optional code EM)”.

4

1

2

3

5

F060306.ai

Figure 6.3.6 “Pulse Output” Screen

Table 6.3.4 “Pulse Output” Screen

Item

No.

Item Name

Description

1

Connection diagram

Shows the connection diagram on another screen. The diagram is any of

those in

Figure 6.3.9 or Figure 6.3.13 depending on the case.

2

Setting Value

Selects and shows the setting pulse value.

3

Measured (pps)

Shows the measured value. Enter the measured value manually.

4

Output (*1)

in Output Selection

Outputs the pulse.

By clicking “Output” after selecting “Pulse”, the pulse with the value of

“Setting value (2)” is transferred to AXF.

(Only when Verifi cation Tool is in “online” mode)

5

Exit

In Output Selection

Stops pulse output and enables other operations.

(Only when Verifi cation Tool is in “online” mode)

*1: When the AXF protect function is set to “Protect”,

Figure 3.2.2 is displayed after clicking “Output”. Follow 3.2.2.