IM 11M12A01-05E

8-11

8. Detailed Data Setting

8.3.3

Alarm Setting Procedure

Set the alarm setpoints following Table 8.6 listing parameter codes.

Table 8.6 Alarm Parameter Codes

T8.3.3.1.EPS

Set item

Parameter code

Set value

Oxygen concentration high-high alarm setpoint

D01

0-100% O

2

Oxygen concentration high-alarm setpoint

D02

0-100% O

2

Oxygen concentration low-alarm setpoint

D03

0-100% O

2

Oxygen concentration low-low-alarm setpoint

D04

0-100% O

2

Humidity high-high alarm setpoint

D05

0-100% H

2

O

Humidity high-alarm setpoint

D06

0-100% H

2

O

Humidity low-alarm setpoint

D07

0-100% H

2

O

Humidity low-low-alarm setpoint

D08

0-100% H

2

O

Mixing ratio high-high alarm setpoint

D11

0-1 kg/kg

Mixing ratio high-alarm setpoint

D12

0-1 kg/kg

Mixing ratio low-alarm setpoint

D13

0-1 kg/kg

Mixing ratio low-low-alarm setpoint

D14

0-1 kg/kg

Oxygen concentration alarm hysteresis

D30

0-9.9% O

2

Humidity hysteresis

D31

0-9.9% H

2

O

Mixing ratio hysteresis

D32

0-0.1 kg/kg

Delay alarm action

D33

0-255 seconds

Oxygen concentration high-high alarm detection

D41

0: Not detected
1: Detected

Oxygen concentration high-alarm detection

D42

0: Not detected
1: Detected

Oxygen concentration low-alarm detection

D43

0: Not detected
1: Detected

Oxygen concentration low-low-alarm detection

D44

0: Not detected
1: Detected

Humidity high-high alarm detection

D45

0: Not detected
1: Detected

Humidity high-alarm detection

D46

0: Not detected
1: Detected

Humidity low-alarm detection

D47

0: Not detected
1: Detected

Humidity low-low-alarm detection

D48

0: Not detected
1: Detected

Mixing ratio high-high alarm detection

D51

0: Not detected
1: Detected

Mixing ratio high-alarm detection

D52

0: Not detected
1: Detected

Mixing ratio low-alarm detection

D53

0: Note detected
1: Detected

Mixing ratio low-low-alarm detection

D54

0: Not detected
1: Detected

CAUTION

Even with alarms set, if “Not detected” has been set in the above alarm detection, no
alarm is issued. Be sure to set “Detected” in the above alarm detection if you use alarm
features.