Table 10.6 Resetting

Switch operation Display Brief description

.

`

ENT

Err-01

If an error occurs, the error number and "------" are displayed

/------

alternately, as given on the left.

.

`

ENT

PASSno

Hold down the [ENT] key for at least three seconds.

.

`

ENT

0

000

Touch the [ENT] key again to switch to the password entry display.

.

`

ENT

1

000

Enter the password 1102.

Intermediate switch operations omitted.

.

`

ENT

1102

.

`

ENT

A

01

.

`

ENT

G

01

Change the parameter code to "G30."

.

`

ENT

G

0

1

.

`

ENT

G

3

0

.

`

ENT All the digits Touch the [ENT] key to execute resetting.

light up.

T10.5aE.EPS

The symbol ( ) indicates that the key are light characters indicate "flashing".

"/" indicate that the characters are displayed alternately.