11. Inspection and Maintenance

11.3 Replacement of Flowmeter for ZR20H Autocalibration

Unit

(1) Remove pipe holding piping connection.
(2) Remove bolts holding flowmeter, and replace it. A white back plate (to make the

float easy to see) is attached. The end of the pin holding down the back plate must
be on the bracket side.

(3) Replace piping, and fix M6 bolts between brackets. *1

*1 : When disassembling and reassembling, mark original positions, and tighten an

extra 5-108 when reassembling. After tightening, do a liquid leakage test.

C

C

B

B

C’

B’

A’

A

A

Zr20h_g0.eps

Vertical mounting

Horizontal mounting

Fixing screw pairs

Connect piping pairs A-A’, B-B’, C-C’

Figure 11.9 Fixing Flowmeter