When setting calibration timing requirements, bear the following precautions in mind:

CAUTION

(1) If the calibration interval is shorter than the sum of stabilization time plus calibration

time, the second calibration start time will conflict with the first calibration. In such a
case, the second calibration will not be conducted. (When both zero and span
calibrations are to be performed, the calibration time is double that required for a
single (zero or span) calibration.)

(2) For the same reason, if the calibration start time conflicts with manual calibration or

semi-automatic calibration, the current calibration will not be conducted.

(3) If the calibration time conflicts with maintenance service or blowback operations,

calibration will start after completing the maintenance service or blowback opera-
tions (see Section 8.2.1, earlier in this manual).

(4) If 000 days, 00 hours are set for the calibration intervals, only the first calibration

will be conducted; a second or later calibration will not be conducted.

(5) If a past date is set to the calibration start day, no calibration will be conducted.

Table 9.1 Parameter Codes for Calibration Setting

Set Item

Parameter code Set value

Engineering unit

Zero-gas concentration

B01

Set Zero-gas concentration

%O

2

Span-gas concentration

B02

Set Span-gas concentration

%O

2

Calibration mode

B03

0 : Manual calibration
1 : Semi-automatic and manual
2 : Automatic, semi-automatic, and manual

Output stabilization

B04

0 minutes 0 seconds to 60 minutes 59 seconds

MM.SS

Calibration time

B05

0 minutes 0 seconds to 60 minutes 59 seconds

MM.SS

Calibration period

B06

0 days 0 hours to 255 days 23 hours

Date and time

Start date and time

B07

Date and time of first calibration

YY.MM.DD.HH.MM

Calibration procedure

B08

0 : Zero and span
1 : Span only
2 : Zero only

T9.1.EPS

9.2.1.7Default Values

When the analyzer is delivered, or if data are initialized, the calibration settings are by
default, as shown in Table 9.2.

Table 9.2 Default Settings for Calibration

Item

Default Setting

Calibration mode

Manual

Calibration procedure

Span - zero

Zero-gas (oxygen) concentration

1.00%

Span-gas (oxygen) concentration

21.00%

Output hold (stabilization) time

10 minutes, 00 seconds

Calibration time

10 minutes, 00 seconds

Calibration interval

30 days, 00 hours

Start day and time

00 (YY) 01 (MM) 01(DD) 00:00

T9.2.EPS